In HPLC, how often should I need to run standard with samples? Why? If the retention time is changed with real samples compared to standard, how can I determine if the peak with changed retention time is samples are actually compounds A and B?

Solution Preview

In HPLC analysis, standards are run and their retention times are noted down in order to calibrate the instrument. Essentially this means running the calibration standards at the beginning of the sequence and at the end, and makes the assumption that any changes occurred in a linear manner during the sequence. The data system then changes the calibration incrementally from the beginning to the end, and applies this to the results.

Drift of peaks due to change in retention time is a major issue in HPLC. To identify the causes (shown in blue below) and remedies of these problems needs step-wise logical analysis:

(1) A new column may require several priming injections for the constituents of a sample to irreversibly bind to the most active surface silanol species - each injection being exposed to successively fewer silanol groups and hence producing a slightly different (typically shorter) retention time. To reduce the requirement for priming injections one should choose a high quality stationary phase, which uses Type B (type II) silica and which has a low metal ion content. To speed up the column 'equilibration' process, one should consider injecting a sample at 10x the usual concentration in order to quickly cover the surface silanol ...